Micro Raman Spectrum Measurement System
The micro Raman spectrum measurement system enables visual precision positioning, allowing operators to detect Raman signals from different surface states of a sample while simultaneously displaying the micro-zone status of the detected position on a computer. This greatly facilitates Raman micro-zone detection, making "what you see is what you measure" possible. By integrating an optimized confocal microscopy module with an advanced Raman measurement system, the platform combines precise optical design matched with infinity-corrected microscopic optics, optimized light path arrangements, and flexible structural modular combinations. Additionally, the high-quality lasers and spectrometers provided by CNI offer customers a wider range of options.
Features
- Adjustable laser with stable frequency, avoiding wavelength drift and damage of samples.
- Optimization of the confocal module, ensuring that focal plane of the laser and view focal plane of the objective confocal.
- Detector with high stability, improving the spectral acquisition efficiency and accuracy.
- Modular design, self-assembly, simple operation and easy to carry.
- Wavelength range and resolution can be customized.
Application
- Identify microns impurities of the pills and describe the distribution.
- Distinguish the effective active ingredient and accessories in pills. Describing the distribution and density of occurrence.
- Positioning precisely the boundary of different particles in ore or arts.
- Describing and locating material contained within mineral or rock.
- The identification of trace material evidence, such as the drug, medicine, explosives, textile and so on
Basic components
Product Model : Mic-R-Sys405 / 488 / 532 / 633 / 785
Basic components : Raman module, Raman detection system, microscope
Product Specifications
Microscope
Microscope model | GMM-580 |
Z-axis focusing system | Coarse and fine Coaxial focus |
X/Y Mobile stages size | 210mm*140mm |
Stage moving range | 75mm*50mm |
Camera | 5 Megapixel Imaging System GMW500 |
Microscope objective | 5X-100X (Optional) |
Camera
Chip | CMOS, colorized, progressive scanning |
Chip size | 1/2.5 inch |
Effective pixels | 5 million |
Pixels size | 2.2μm*2.2μm |
Max Resolution (H×V) | 2592*1944 |
Frame rate | 5fps@2592 ×1944 |
Exposure time | 0.21~2000ms |
Noise of amplitude | 40.5dB |
Operating temperature(°C) | -10 ~ 50 |
Data interface | High speed USB2.0 |
Lens mount | Standard C-mount |
Laser device
Wavelength (nm) | 405±0.5 | 488±0.5 | 532±0.5 | 633±0.5 | 785±0.3 | 830±5 |
Mode | Single | Single | Single | Single | Single | Multi- |
Spectral line width (cm-1) | 1.8 | 3.5/ 1.2 | 3.5/ 0.1 | <0.7 | <3.2 | <28 |
Output power (mW) | 1-500 | 1-150 | 1-1000 | 1-80 | 1-500 | 1-900 |
Power stability (pk-pk, 4 hours) | <1%,<3%,<5% | <1%,<3%,<5% | <1%,<3%,<5% | <1%,<3%,<5% | <1%,<3%,<5% | <1%,<3%,<5% |
Parameter of fiber | 100um@0.22NA | 100um@0.22NA | 100um@0.22NA | 100um@0.22NA | 100um@0.22NA | 100um@0.22NA |
Noise of amplitude(rms) | - | <1% | <0.5% | <1% | <0.5%,<1% | <1% |
Operating mode | CW | CW | CW | CW | CW | CW |
Warm-up time(minutes) | <5 | <5 | <10 | <5 | <5 | <5 |
Operating temperature (℃) | 10-35 | 10-35 | 15-35 | 10-35 | 10-40 | 10-35 |
Expected lifetime (hours) | 10000 | 10000 | 10000 | 10000 | 10000 | 10000 |
Warranty | 1 year | 1 year | 1 year | 1 year | 1 year | 1 year |
Spectrometer
Wavelength range | 488-572nm (488nm excitation) | 532-700nm | 633-781nm (633nm excitation) | 785-1100nm | 830-1000nm |
Optical resolution | 8-12cm-1 (488nm excitation) | 8-12cm-1 | 8-12cm-1 (633nm excitation) | 5-8cm-1 | 5-7cm-1 |
Pixels | 2048 pixels | ||||
Pixels size | 14um*14um | ||||
Dimensions (mm) | 149*109*50 | ||||
Weight (g) | 1000 | ||||
Signal-to-noise ratio | 1000:1 | ||||
Integration time | 17ms-10s | ||||
Power consumption | 450mA@5VDC | ||||
Data transfer speed | Full scans into memory every 4 milliseconds with USB 2.0 | ||||
Connector | 30pin connector | ||||
Raman probe
Excitation wavelength (nm) | 488 | 532 | 633 | 785 | 830 |
Probe focus length (mm) | 7.5 | ||||
Probe main material | Probe material is 2A12 after Blackening; Stainless steel detector tip | ||||
Fiber configuration | Two single fiber (100um excitation fiber, 200um optical fiber for collecting light); N.A.=0.22, PVC/metal sheath; 1m Length | ||||
Software
Optical spectrum analysis | Spectral Analysis / Raman Analysis 2.0 |
Operating system | Windows XP/ Win7/ Win10 (32bits or 64 bits) |