Portable Raman Spectrometer


CNI has developed portable Raman spectrometers—RamanSys-405, RamanSys-532, and RamanSys-785, with features of portability, high resolution, and high sensitivity, these systems are suitable for applications such as food safety, environmental science, and judicial identification. 

 

 

Features

  • Spectrometer Resolution
  • Broad spectral range, Raman drift range
    RamanSys-405: <12 cm -1 ; RamanSys-488: <12 cm -1 
    RamanSys-532: <12 cm -1 ; RamanSys-633: <12 cm -1 
    RamanSys-785: <5 cm -1 ; RamanSys-830: <7 cm -1
  • Broad spectral range, Raman drift range
    RamanSys-405: 300–3000 cm⁻¹ -1 ; RamanSys-488: 450–3000 cm⁻¹ -1 
    RamanSys-532: 150–4500 cm⁻¹ -1 ; RamanSys-633: 110–3000 cm -1 
    RamanSys-785: 150–3648 cm⁻¹ -1 ; RamanSys-830: 200–2048 cm -1
  • Customized wavelength range and resolution
  • Fit various applications
  • Compatible with software Raman Analysis 2.0

Application

  • Raman spectral analysis
  • Chemical or biological research
  • Environmental sciences
  • Diamond identification
  • Food safety testing
  • Judicial identification
  • Pharmaceutical and medical diagnosis
  • Geological exploration
  • Hazard detection
 

 

Basic components

Portable Raman Spectrometer

405nm/ 488nm/ 532nm/ 633nm/ 785nm/ 830nm fiber coupled laser

Spectrometer: Raman Standard: Sunshine GE-Raman

RPB-405 / 488 / 532 / 633 / 785 / 830 Raman Probe

PC Data Processing Unit 
Man-machine interface

 

 

Specifications

 

Laser

Length (nm)

405±0.5

488±0.5

532±0.5

633±0.5

785±0.3

830±5

Mode

Single

Single

Single

Single

Single

Multi-

Spectral line width (cm-1)

1.8

3.5/ 1.2

3.5/ 0.1

<0.7

<3.2

<28

Output power (mW)

1-500

1-150

1-1000

1-80

1-500

1-900

Power stability (pk-pk, 4 hours)

<1%,<3%,<5%

<1%,<3%,<5%

<1%,<3%,<5%

<1%,<3%,<5%

<1%,<3%,<5%

<1%,<3%,<5%

Parameter of fiber

100um@0.22NA

100um@0.22NA

100um@0.22NA

100um@0.22NA

100um@0.22NA

100um@0.22NA

Noise of amplitude(rms)

-

<1%

<0.5%

<1%

<0.5%,<1%

<1%

Operating mode

CW

CW

CW

CW

CW

CW

Warm-up time(minutes)

<5

<5

<10

<5

<5

<5

Operating temperature (℃)

10-35

10-35

15-35

10-35

10-40

10-35

Expected lifetime (hours)

10000

10000

10000

10000

10000

10000

Warranty

1 year

1 year

1 year

1 year

1 year

1 year

Spectrometer

Wavelength range

405-461nm
(405nm excitation)

488-572nm
(488nm excitation)

532-700nm
(532nm excitation)

633-781nm
(633nm excitation)

785-1100nm
(785nm excitation)

830-1000nm
(830nm excitation)

Optical resolution

8-12cm-1
(405nm excitation)

8-12cm-1
(488nm excitation)

8-12cm-1
(532nm excitation)

8-12cm-1
(633nm excitation)

5-8cm-1
(785nm excitation)

5-7cm-1
(830nm excitation)

Pixels

2048 pixels

Pixels size

14um*14um

Dimensions (mm)

149*109*50

Weight (g)

1000

Signal-to-noise ratio

1000:1

Integration time

17ms-10s

Power consumption

450mA@5VDC

Data transfer speed

Full scans into memory every 4 milliseconds with USB 2.0
Full scans into memory every 18 milliseconds with USB 1.1

Connector

30pin connector

  Raman probe

Excitation wavelength (nm)

405488

532

633

785±

830

Probe focus length (mm)

7.5

Probe main material

Probe material is 2A12 after Blackening; Stainless steel detector tip

Fiber configuration

Two single fiber (100um excitation fiber, 200um optical fiber for collecting light); N.A.=0.22, PVC/metal sheath; 1m Length

 

Software

Optical spectrum analysis

Spectral Analysis / Raman Analysis 2.0

Operating system

Windows XP/ Win7/ Win10 (32bits or 64 bits)

 

 

Characteristic

  • Dimensions: RamanSys-405/ 785:340*295*152mm; RamanSys-532:418*330*173mm
  • Adjustable and stable 405/ 488/ 532/ 633/ 785/ 830nm laser source
  • Simple, fast measurement and easy operation
  • Function with matching library and searching, provided convenient self-built library function
  • Flexible and stable sampling, various forms in analyzing samples
  • External power supply (charging), with vehicle-mounted charger
  • Dustproof, waterproof, anti-vibration, Anti-glare interference in wild environment usage
 

 

Application

Pharmaceuticals and Medical Diagnostics

Jewelry identification

Medicine identification

Food and Agriculture

Chemical or biological research

Food Safety Testing

Textile Printing and Dyeing Industry

Hazard detection